AU recall: Philips BiPAP A40 – non-conforming materials could cause off-gassing or device failure

A recall has been issued in Australia for Philips bi-level positive airway pressure (BiPAP) A40 non-invasive ventilators with the following serial numbers:

  • V2887246677A0
  • V288724698F57
  • V28872472285C
  • V288724777FF1
  • V28872479968F
  • V288724808886
  • V28872485DF2B
  • V28872492B24C
  • V28872495C6F3
  • V288724981D16
  • V288725010D13
  • V2887251114CB
  • V2887251771FD


These units may have been manufactured with non-conforming plastic materials in the motor assembly. Potential hazards from the non-conforming materials are:

  • biosafety or toxicological hazard from off-gassing.
  • sudden device failure, which could lead to patient asphyxia if not immediately identified and addressed.


Philips advises it will replace all affected devices with a new A40 Pro unit.

For more information, see the Therapeutic Goods Administration Australia recall listing.

At the time of writing, no recall is listed with Medsafe New Zealand.

Last updated 12:52 9.09.2022